07

PHscan20

Cal

Enter

PHscan

POCKET PH TESTER

allows up to 3 points calibration.

tester can

 

 

   

 

 

 

Press CAL key the tester shows “pH7.00/CAL1”(or “pH6.86/CAL1”).

4. Dip the

 

pH electrode into the

 

pH7.00

 

(or

 

pH6.86)

 

standard

 

buffer

 

solution, 

    the end 

 

of 

 

the electrode must be completely immersed into

 

the

 

calibration

    solution. Stir the tester gently.

5. Press 

 

ENTER 

 

key 

 

to confirm. W  

 

 

 

 

   

 

    

pH Calibration

During pH mode, the tester 

       

 

 We recommend

that you perform

 

at

 

least 2

 

points calibration for high accuracy

 

measurement.

The 

Single point calibration should only be carried out with pH 7.00 or pH 6.86, 

   

 

otherwise calibration will not be accepted.

 

 automatically recognise and calibrate to following standard

buffer values.

USA Standard: 

 

pH4.01, 7.00, 10.01

NIST Standard: pH4.01, 6.86, 9.18

The

 

tester

 

must be calibrated prior to first use or new electrode replaced. To

ensure

 

accuracy,

 

regular

 

calibration

 

is

 

recommended.

 

Do

 

not

 

reuse

 

calibration

solution

 

after

 

calibration,

 

contaminants

 

in

 

solution

 

will

 

affect

 

the

 

calibration

 

and

eventually the accuracy of the measurement.

Single Point Calibration

1. Make sure that you have selected 1 point calibration in the system menu.
2. Rinse the pH

 

electrode

 

with distilled water, press MEAS

 

key

 

to

 

turn

 

on

 

the

    tester.
3. 

 

ait for the measured value to stabilize, the

display shows “END”. Single point calibration is done.

PHscan20

08

Enter

Enter

2 Points Calibration

Make sure that you have selected 

 

2 points calibration in the system menu.

Repeat

 

steps

 

2

 

to

 

5

 

above.

 

When

 

the

 

first

 

calibration

 

point

 

is

 

completed

,

 

the

    display will show 

 

“CAL2”. The tester prompts you to continue with second

    point calibration.

Rinse 

 

the 

 

electrode 

 

with 

 

distilled 

 

water, 

 

ait for the measured value to stabilize, the

    display shows electrode slope and “END”. Second point calibration is done.

3 Points Calibration

Make sure that you have selected 

 

3 points calibration in the system menu.

Repeat

 

steps

 

2

 

to

 

5

 

above.

 

When

 

the

 

first

 

calibration

 

point

 

is

 

completed

,

 

the

    display will show

 

“pH4.01/CAL2”.

 

The tester prompts you to continue with 

    second point calibration.

Rinse 

 

the 

 

electrode 

 

with 

 

distilled 

 

water, 

 

ait for the measured value to stabilize, the

    display shows electrode slope and “pH10.01/CAL3” (or “pH9.18/CAL3”).

 

 

 

 

 

 

dip the electrode into the pH4.01

    or 10.01

 

(pH4.01 or 9.18) standard buffer solution. 

 

 

Press 

 

ENTER 

 

key 

 

to 

 

confirm. 

 

W  

 

 

 

 

   

 

 

 

 

 

 

 

 

 

 

 

dip the electrode into the pH4.01

    standard buffer solution. 

 

 

Press 

 

ENTER 

 

key 

 

to 

 

confirm. 

 

W  

 

 

 

 

   

 

Summary of Contents for PHscan20

Page 1: ...come our customers suggestions on how to speed up these efforts Hazardous Substance Statement Warranty The warranty period for PHscan series tester is 12 months from the date of shipment Above warrant...

Page 2: ...ER HOLD DESCRIPTION Indicates tester is in the measurement mode Indicates tester is in the calibration mode Indicates tester is in SETUP mode Indicates the measured value has been frozen Indicates tem...

Page 3: ...or 8 minutes the tester will automatically turn off to conserve energy To disable the auto off function please read the SETUP MENU section The tester contains an integrated system menu that allows you...

Page 4: ...ress ENTER key to confirm the tester shows YES OFF auto off function is enabled points calibration indicating that indicating that the Enter Cal When the auto off function is enabled 8 Press CAL key t...

Page 5: ...e selected 1 point calibration in the system menu 2 Rinse the pH electrode with distilled water press MEAS key to turn on the tester 3 ait for the measured value to stabilize the display shows END Sin...

Page 6: ...the measured reading you can continue to take measurement Electrode Care and Maintenance For best results please always keep the sensor s glass bulb wet Ensure that the sensor is thoroughly washed wi...

Page 7: ...eparation is done ORDER CODE ITEM E PHSCAN ST Standard pH Electrode Used for routine pH measurements e g drinking water rainwater etc E PHSCAN FT Flat Surface pH Electrode Used for measuring the skin...

Reviews: