PHscan20

Enter

Rinse

 

the

 

electrode

 

with

 

distilled

 

water,

 

 

 

 

 

 

ait for the measured value to stabilize, the

    display shows electrode slope and “END”. The tester returns to the pH
    measurement mode. Calibration is completed.

    EXIT THE CALIBRATION:
    

During

 

the

 

calibration,

 

if

 

you

 

want

 

to

 

exit

 

calibration

 

mode,

 

press

 

MEAS

 

key.

    The tester will immediately return to measurement mode.

 

 

dip the electrode into the pH10.01

    (or pH9.18) standard buffer solution. 

 

 

Press 

 

ENTER 

 

key 

 

to 

 

confirm. 

 

W  

 

 

 

 

   

 

pH Measurement

Rinse

 

the

 

electrode

 

thoroughly

 

with

 

distilled

 

water,

 

 

 

   

ait for the

to stabilize, record

    

 

1. 

press MEAS key to turn

    on the tester.
2. Dip the electrode into the sample solution, stir the tester gently.
3. W  

 

 reading   

 the measured value on the display.

Auto-Hold Function

When 

 

the auto-hold 

 

function is enabled, 

 

the 

 

tester 

 

will 

 

automatically sense a

stable

 

end-point

 

measured

 

value

 

and

 

freeze it,

 

“HOLD”

 

indicator will appear on

the display. Press HOLD key, the tester resumes measuring.

Manual Hold Function

1. During 

 

the measurement, press HOLD key, 

 

,

 

“HOLD”

 

indicator

 

will

 

appear

 

on the display.

2. Press key again to

 

release the measured reading,

 

you

 

can

 

continue to take

    measurement.

Electrode Care and Maintenance

  For best results, please always keep the sensor’s glass bulb wet.

  

Ensure

 

that

 

the

 

sensor

 

is

 

thoroughly

 

washed

 

with

 

distilled

 

water

 

after

 

each

use.

the tester immediately freezes

    currently displayed value

PHscan20

10

Hold

Meas

  If 

 

you do not use 

 

the tester for long periods, please store the sensor with

electrode storage solution. DO NOT use de-ionized or distilled water.

    

 

Electrode Replacement

When 

 

the 

 

tester fails 

 

to calibrate or gives 

 

fluctuating readings for calibration 

standards, you need to replace the electrode module.
1. Twist the electrode collar counter clockwise, 

 

pull the old electrode module

    away from the tester.

2. Align the slot on the new electrode

 

module,

 

gently

 

push

 

the

 

module

 

into

 

the

    tester.

3. Twist the electrode

 

collar

 

clockwise until it is tight.

 

Troubleshooting

Installation

 

is

 

completed.

Slot

 Electrode dried out.

   

 

Soak the

 

electrode 

 

in

 

tap

 

water

 

for

 

10

 

minutes. 

 Measured value is out of range.

   

 

Check the electrode whether clogged, dirty or broken.

module

 Incorrect buffer used or contaminated buffer solution.

   

 

Use fresh pH buffer solutions for calibration.

 Electrode is broken.

   

 

Replace the electrode 

.

module

LCD DISPLAY

CAUSE

 Keypad is not working properly.

 

  

Replace the batteries.

Summary of Contents for PHscan20

Page 1: ...come our customers suggestions on how to speed up these efforts Hazardous Substance Statement Warranty The warranty period for PHscan series tester is 12 months from the date of shipment Above warrant...

Page 2: ...ER HOLD DESCRIPTION Indicates tester is in the measurement mode Indicates tester is in the calibration mode Indicates tester is in SETUP mode Indicates the measured value has been frozen Indicates tem...

Page 3: ...or 8 minutes the tester will automatically turn off to conserve energy To disable the auto off function please read the SETUP MENU section The tester contains an integrated system menu that allows you...

Page 4: ...ress ENTER key to confirm the tester shows YES OFF auto off function is enabled points calibration indicating that indicating that the Enter Cal When the auto off function is enabled 8 Press CAL key t...

Page 5: ...e selected 1 point calibration in the system menu 2 Rinse the pH electrode with distilled water press MEAS key to turn on the tester 3 ait for the measured value to stabilize the display shows END Sin...

Page 6: ...the measured reading you can continue to take measurement Electrode Care and Maintenance For best results please always keep the sensor s glass bulb wet Ensure that the sensor is thoroughly washed wi...

Page 7: ...eparation is done ORDER CODE ITEM E PHSCAN ST Standard pH Electrode Used for routine pH measurements e g drinking water rainwater etc E PHSCAN FT Flat Surface pH Electrode Used for measuring the skin...

Reviews: