TDSscan20   

I

    4 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
     
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 
5. 

Press the 

Cal

 key to modify the temperature value. 

 

 

 

 
6. 

Press the 

Enter

 key to save and press the 

Meas

 key to return 

 

to the measurement mode. 

 

 

 

 

 

 

Measurement 

 
Rinse the electrode with distilled water, place the electrode into the 
sample solution and stir gently, make sure that no air bubbles on the 
sensor surface. Wait for the measurement to stabilize and record the 
reading. 
 

 

 

During the measurement, DO NOT completely immerse the tester 

 

in water. 

 

 

 

 

If the display shows ---- indicating the measurement exceeds the 

 

range, remove the tester from the sample immediately. 

 

 

If the

 

 

option is enabled in the setup menu, the tester will 

 

automatically lock the measurement endpoint and show

 

HOLD 

 

icon. Press the 

Meas

 key to resume measuring. 

 

 

 

 

 

 

Electrode   
Maintenance and Replacement 

 

Rinse the electrode thoroughly with distilled water after use. 

 

Do not touch the platinum black coating on the sensor surface 

 

and always keep it clean. 

 

If there is a build-up of solids inside the sensor, remove carefully, 

 

then recalibrate the tester. 

 

If you do not use the tester for long periods, store the electrode 

 

with tap water. 

 

Replacing the Electrode 

If the tester fails to calibrate or gives fluctuating readings, you should 
consider replacing the electrode. 
1. 

Twist the electrode collar counter clockwise, pull the electrode 

 

away from the tester. 

 

 

2. 

Align the slot on the new electrode, gently push the electrode 

 

into the tester. 

 

 

3. 

Twist the electrode collar clockwise until tight. 

 

 

 