14. Maintenance
Max
Min
1
2
Image 14-8
Cooling liquid check
1
Max level
2
Min level
If the current liquid cooling level is somewhere between
Min
and
Max
, then no action should be taken.
If the current liquid cooling level is approaching the
Min
level point, than check for leakage in the cooling system prior starting
the re
fi
ll procedure.
If you cannot see the level, open the
fi
ller cap and shine a torch into the reservoir
14.9 Cooling liquid re
fi
ll
W
ARNING
:
This procedure may only be performed by quali
fi
ed technical service personnel.
Order info cooling liquid
B1909086K : bottle of 1 liter cooling liquid.
Necessary parts
Cooling liquid
How to re
fi
ll
1. Remove the side cover, "Removal of the side cover", page 126.
2. Open the
fi
ller cap by turning it anti clockwise.
R5906753 DPXK BLP SERIES 20/11/2017