7. Coolant procedures
7.6
Cooling liquid visual check
Ensure the chiller is turned off before connecting the open-end hose to the bypass tube. If not, spillage will
occur.
How to perform
1. Connect the bypass tube to the main cooling circuit quick couplers as illustrated.
Image 7-24
Connecting service hose to lower quick coupler
2. Insert the open end of an open ended hose (included in the service kit) into a container and connect the other end to the bypass
tube.
Image 7-25
Note:
Ensure you
fi
rst place the open-ended hose in the container before connecting it to the bypass tube. If not, there is a big
chance you will spill
fl
uid.
3. Switch on the power.
The chiller will notice a pressure difference due to the open-ended hose and the pump will activate. As a result, cooling
fl
uid will
fi
ll the container.
4. Switch off the power after collecting about 1l of cooling liquid.
Note:
If the power is not switched off, the pressure pump will continue to pump cooling liquid in the tank in bursts of 30 seconds.
5. Perform a visual check. The cooling liquid should be blue (azure/cerulean) with only minor levels of particles being present.
-
If the cooling liquid is okay, add the coolant back to the tank. See procedure "Adding coolant to the tank", page 31.
-
If the cooling liquid shows unacceptably high levels of contamination or has changed color, the cooling liquid must be replaced
entirely. See procedure "Draining the coolant circuit", page 26.
R5906015 ULC-30A 01/02/2018
35
Summary of Contents for ULC-30A
Page 1: ...ULC 30A Service manual R5906015 01 01 02 2018...
Page 4: ......
Page 6: ...Table of contents Index 97 2 R5906015 ULC 30A 01 02 2018...
Page 10: ...1 Safety 6 R5906015 ULC 30A 01 02 2018...
Page 24: ...5 Covers 20 R5906015 ULC 30A 01 02 2018...
Page 28: ...6 Fan unit 24 R5906015 ULC 30A 01 02 2018...
Page 40: ...7 Coolant procedures 36 R5906015 ULC 30A 01 02 2018...
Page 70: ...10 Refrigeration unit 66 R5906015 ULC 30A 01 02 2018...
Page 92: ...11 Electronics Unit 88 R5906015 ULC 30A 01 02 2018...