11. Electronics Unit
11.5.3 Removing the V1000 pump inverter
Necessary tools
•
Phillips screwdriver PH1
•
Allen wrench 3 mm
•
Small
fl
at screwdriver
This procedure assumes that power has been switched off. Unless otherwise stated, power must be off prior
to start with this procedure. To switch off power, see procedure "Switching off power and locking the switch",
page 15.
How to remove the pump inverter?
1. Remove the front cover. See procedure "Removing a cover", page 17.
2. After switching off power, wait for at least 1 minute to allow the inverter capacitor to discharge.
3. Unplug the cable connector at the front cover of the pump inverter.
Image 11-39
Unplugging the cable connector
4. Remove the screw (reference 1, image 11-40) on the front cover of the pump inverter. Use a PH1 Phillips screwdriver.
1
Image 11-40
Removing the screw
5. Remove the cover of the inverter.
82
R5906015 ULC-30A 01/02/2018