6
The StatLock® stabilization device may be included in PowerPICC® Provena™ Catheter kits. Please refer to Statlock® stabilization
device Instructions For Use on the proper use and removal. The StatLock® stabilization device should be monitored daily and
replaced at least every seven days.
The StatLock® Stabilization Device Procedure
Single Lumen
1. Secure catheter with
StatLock® stabilization device.
2. Cover site and StatLock®
stabilization device with
transparent dressing.
3. Place anchor tape sticky side
up, under hub. Wedge tape
between hub and wings.
4. Chevron anchor tape on top
of transparent dressing.
Dual Lumen
1. Secure catheter with
StatLock® stabilization device.
2. Cover site and StatLock®
stabilization device with
transparent dressing.
3. Place 1st anchor tape sticky
side up, under one extension
leg. Wedge tape between hub
and wings. Chevron anchor
tape on top of transparent
dressing.
4. Place 2nd anchor tape sticky
side up under remaining hub.
Wedge tape between hub and
wings. Chevron anchor tape on
top of transparent dressing.
Tape Strip Stabilization Procedure
Single Lumen
1. Place 1st anchor tape over
wings.
2. Cover site and 1st anchor
tape with transparent dressing
up to hub, but not over hub.
3. Place 2nd anchor tape sticky
side up, under hub. Wedge tape
between hub and wings.
4. Chevron anchor tape on top
of transparent dressing.
Dual Lumen
1. Place 1st anchor tape over
wings.
2. Cover site and 1st anchor
tape with transparent dressing
up to hub, but not over hub.
3. Place 2nd anchor tape sticky
side up, under one extension
leg. Wedge tape between hub
and wings. Chevron anchor
tape on top of transparent
dressing.
4. Place 3rd anchor tape sticky
side up under remaining hub.
Wedge tape between hub and
wings. Chevron anchor tape on
top of transparent dressing.
14. Verify Placement
A.
PICCs should be positioned with the catheter tip in the lower 1/3 of the SVC. Verify correct catheter tip position using
radiography or other appropriate technology.
15. Power Injection Procedure
WARNING: PowerPICC® Provena™ Catheter indication for power injection of contrast media implies the catheter’s ability to
with stand the procedure, but does not imply appropriateness of the procedure for a particular patient. A suitably trained
clinician is responsible for evaluating the health status of a patient as it pertains to a power injection procedure.
A.
Remove the injection/needleless/end cap from the PowerPICC® Provena™ Catheter.
B.
Attach a 10 mL or larger syringe filled with sterile saline.