Operation
Preferences may be set up for the various programs:
The heating duration can be preset for the "PLASMA", "BLOOD" and "HPC" pro-
grams.
With the "PLASMA" program, the nominal temperature of the heat transfer fluid can
be preset in the service menu. Details are described in a separate service description
which can be requested from Barkey GmbH & Co. KG.
Other parameters for the "USER" program can also be preset in the service menu.
Details are described in a separate service description which can be requested from
Barkey GmbH & Co. KG.
Changing the remaining time for the "PLASMA", "BLOOD", "HPC" and "USER" pro-
grams with the left-hand (" -") or right-hand selection button (" +") is described in the
Chapter Changing the remaining time.
If the confirm button ("Stop") is pressed while a program is running, the program is can-
celled and the main screen is displayed again (see the Chapter Messages after the end of a
program).
NOTE
Since no remaining time is active with the "CONTINUOUS" program, the display
shows "--:--". Changing the remaining time is not possible.
4.4.2
Changing the remaining time
While a program is running, the user can alter the remaining time – usually between 30s
and 99min – with the left or right-hand selection button (" -" or " +"). If the current
remaining time is already less than 30s, then this counts as the minimum permitted input
so that the remaining time cannot be unintentionally increased.
NOTE
Since no remaining time is active with the "CONTINUOUS" program, the display
shows "--:--". Changing the remaining time is not possible.
If the current remaining time can be changed, it is initially frozen for the current program
when one of the two selection buttons (" -" or " +") is pressed. Depending on which
button is pressed, the displayed remaining time is then rounded up or down to the nearest
whole 10s and the display flashes to optically highlight the active remaining time adjust-
ment.
After the remaining time has been changed, pressing the confirm button ("OK") enters the
new setting, and the current program is resumed.
If neither one of the two selection buttons (" -" or " +") nor the confirm button ("OK")
is pressed within 5s, any new remaining time that has been selected will be rejected, the
original remaining time is reactivated and the current program is resumed.