14
2. Select the PROCESS - HI screen by pressing the MENU key. The follow-
ing screen will appear (refer to Figure 17). Set the Process HI tempera-
ture by using the arrow keys.
PROCESS ALARM
HI XXXXX °F
FIGURE 17. PROCESS HI SETTINGS SCREEN
Process LO
1. Process LO is a fixed value and will not change if the SP temperature
value is changed. It activates the alarm when the PV temperature drops
below the alarm temperature. Refer to Figure 18. If the Control Action is
set to HEAT and the PV temperature is below the alarm temperature
when the RUN key is pressed, the alarm will not be activated. The PV
temperature must first rise above the alarm setpoint before the low alarm
is enabled. This allows for normal system startup without activating the
alarm.
ALARM MODE
PROCESS - LO
FIGURE 18. PROCESS LO SCREEN
2. Select the PROCESS - LO screen by pressing the MENU key. The fol-
lowing screen will appear (refer to Figure 19). Set the Process LO tem-
perature by using the arrow keys.
PROCESS ALARM
LO 000XX °F
FIGURE 19. PROCESS LO SETTINGS SCREEN