29
10. Allow the system to operate normally until
the reservoir is full and the unit goes into
standby.
11. Allow the solution to sit in the system for
30-45 minutes.
12. Reconnect the quick disconnect insert(s)
and product water tubing removed in step
4. Direct the tubing to drain and drain the
storage reservoir. After the reservoir is
empty, disconnect the quick disconnect
insert(s) and product water tubing again.
13. Remove the top and bottom membrane
bypass tubing assemblies and retain them
for future use. Remove the sanitization
cartridge and discard it.
14. Install a new pretreatment cartridge.
Reinstall existing membrane cartridges if
they were operating properly.
15. Reset pretreatment timer. (See “Prefilter
Replacement.”)
16. Operate normally. Drain the first tankful of
water as described in step 12.
17. Reconnect the product water tubing to the
EASYpure unit. Reconnect the product
water tubing to the NANOpure. Reconnect
the EASYpure unit or NANOpure to the
power supply.
M
AINTENANCE
Note
Reinstalling your old mem-
branes may reintroduce bac-
teria into your EASYpure RO
system.