6
Specifications
Dimensions
30.5 cm W x 46.7 cm W x 46 cm H.
Feedwater Requirements1, 5
Max. Temperature ........................48.8°C
Min. Pressure (standard unit) ......30 psig
(2.1 bar)
Min. Pressure (low pressure unit) 9 psig
(0.62 bar)
Max. Pressure (standard unit)......100 psig
(6.8 bar)
Max. Pressure (low pressure unit)
15 psig (1.1 bar)
Inlet Flow Requirements ..............53.5 lph
pH ................................................4 - 11
TDS (Max. ppm as CaCO3) ........500
Silt Density Index ........................<5%
Free Chlorine................................<0.1 ppm2
Langlier Saturation Index ............Negative
Turbidity........................................<1 NTU
Iron (Total as Fe) ..........................<0.5 ppm
Product Water
Specifications5
Product Water Flow Rate ............12 lph3
% Recovery ..................................20%-26%
1 Feedwater suitability must be determined by a water analysis.
2 Free chlorine removed by pretreatment cartridge.
3 ±1.5 lph, feedwater at 25°C and 60 psig pressure.
4 Membrane performance is dependent on condition, pressure, recovery, water temperature and water composition. Listed membrane
performance for a new membrane is based on optimum operating conditions of 60 psig, 15% recovery, 25°C (77°F) feedwater tempera-
ture and feedwater composition of a maximum of 500 ppm NaCl at a pH of 6.0-6.5.
5 Membrane performance, flowrates and recovery information are nominal values. Variations in the feedwater temperature, pumps and
reject orifice may slightly alter these results.
Membrane Rejection Rates4
Monovalent Ions ..........................90-95%
Polyvalent Ions ............................95-99%
Particles........................................>99%
Microorganisms ............................>99%
Dissolved Organics (300 MW) ....>99%
Reject Water Flow Rate
650 ml/min
at
60 PSIG
±5
PSIG
Electrical Requirements
Model D7422-33 - 230 VAC ±10%, 47-63 Hz
Model D7429-33 - 230 VAC ±10%, 47-63 Hz
Environmental Conditions
Operating: 17°C - 27°C; 20% - 80% relative
humidity, non-condensing. Installation
Category II (over-voltage) in
accordance with IEC 664. Pollution
Degree 2 in accordance with IEC
664. Altitude limit: 2,000 meters.
Storage: -25°C - 65°C; 20% - 80% relative
humidity.