14
Water Connections
Feed Water Connection
1.
Locate the length of 3/8” O.D. tubing provided
with the unit with a quick disconnect insert on
one end and a 3/8” (.95 cm) O.D. X 1/4” (.64
cm) NPT tubing adapter on the other.
2.
Install the tubing adapter onto your
incoming water line. We recommend a
customer supplied shut off valve be
installed in your feed water line. Do not
connect the feed water to your NANOpure
DIamond Life Science (UV/UF). You will connect
the feedwater during the “Initial Sanitization”
procedure explained later in this manual. Note:
The cartridge pack and ultrafilter must be
installed before connecting water line.
3.
If a pressure switch is to be used, see the
“Installing a Float or Pressure Switch” section
for more information.
Atmospheric Drain Connection
When the NANOpure DIamond Life Science (UV/UF)
flushes its membrane, the water used is sent to drain
through this connection. To install:
1.
Locate the drain water tubing. This is the 1/4”
(.64 cm) O.D. tubing that is approximately 6 ft.
(1.9 m.) long. The atmospheric drain fitting is
located on the lower left side of the NANOpure
DIamond Life Science (UV/UF).
2.
Route the other end of the drain water tubing to
an atmospherically vented drain and make a
connection. Ensure there are no kinks in the
tubing and that it proceeds in a downward
plane. Proceed to “Initial Operation.”
I
NSTALLATION
Water Connections - Left side view
Caution
Do not connect feed water until direct-
ed to do so during “Initial Sanitization.”
Pump
Interlock
Remote 2
Remote 1
Feed
Water Inlet
Atmospheric
Drain