25
Control Panel
The DIamond TII is controlled through a panel which
incorporates buttons to control its functions. This control
panel utilizes a graphical display to show system informa-
tion and the purity of the product water. The intensity of
the display may be adjusted according to user preference.
Refer to “Setting System Options.”
Three LED’s located on the control panel will
illuminate to inform you of the system status as follows:
•
Green LED (top center of the display):
Solid ON - system is running
Blinking - system is in STANDBY
•
Red “System Alarm” LED:
Solid ON - Something in the system is causing
an error and must be corrected. To diagnose,
press MENU and select System Alarm.
•
Yellow “Routine Care” LED:
No error exists, but care of product is recom-
mended at this time. To diagnose, press MENU
and select Routine Care.
The control panel can be removed from the unit and
remotely mounted. Please refer to “Installing the Control
Panel in a Remote Location” in the “Installation” section of
this manual for more information about how to remove the
control panel from the unit.
Buttons
When the main power switch is on, the buttons on the
control panel function as follows:
1.
START/STOP
: When the unit is in the IDLE
Mode or STANDBY Mode, pressing the
START/STOP switch will put the DI operation of
the unit in the RUN Mode, turning the DI pump
and UV lamp on. The RO operation will auto-
matically turn on in RUN Mode to fill the reser-
voir as needed. When the unit is in the RUN
Mode pressing the START/STOP switch will put
the DI and RO operation of the unit into the Idle
Mode. When the unit is in any of the flush or
Controls
Figure 18: DIamond TII Control Panel
Note
The removable control panel was not
designed to be repeatedly removed
from the unit, therefore, you may
experience difficulty when attempting
to remove it. It is therefore recom-
mended that the control panel be per-
manently mounted in a remote location
or remain intact with the unit.
Warning
Do not use in the presence of flam-
mable or combustible materials; fire
or explosion may result. This device
contains components which may
ignite such materials.