ULTRASAMPLER® sample collection system
189
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
Installing the bottle drive
Fit bottle drive in refrigerated sample collection box.
Plug connection cable in first if necessary because of space constraints.
Screw down bottle drive with four M6 screws and protect screw fittings.
Loosen branch pipe fasteners. Turn branch pipe with sampler until the
tip of the injection needle is pointing to the centre of the magnetic disc.
Tighten fasteners again.
Removing the bottle drive
Undo the four M6 screws. Remove the bottle drive.
Unplug the connection cable before or after the removal process
depending on space constraints.