Start of operation
199
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
Preparations for the operational check
Prepare 500 l of saltwater for intake.
Concentration: approx. 500 g of salt (1 pack) to 200 l of water.
The saline solution should generate an output voltage of 0.3 ... 0.4 V on the
fill level sensor output when the pipe is full.
It is advisable to put this amount of saltwater through the collection vehicle
twice.
Prepare the collection vehicle for the intake, immersing the intake socket in
the saltwater.
Adjustment and settings on the ULTRASAMPLER system
Once the installation process has been duly completed and all the checks
have been made to ascertain that the maximum ratings of the sampler are
not being exceeded, the electrical settings can be configured.
See configuration instructions 3003_TIGER_MAK_SAK_110625.pdf
Standard settings on the MAK 3003 controller in conjunction with
ULTRASAMPLER controller type 6771-31:
Basic settings
Control mode
Current loop
Flow rate at 20 mA:
1400 l/min
Set pre-sample quantity:
80 l
Pump stop:
85 %
Correction factors:
Low volume range:
0 % (1/X)
.. operative up to:
150 l
Mid volume range:
0 % (1/X)
.. operative up to:
250 l
High volume range:
0 % (1/X)
Start delay with high volume: 2.0 s
Standard rinse:
Yes
Firmware version:
2.34
Sampler 1
Sample mode:
Single sample
Sample bottle volume:
35 ml
Overflow tolerance:
120 %
Start delay:
0.0 s
Calibration factor:
270 µL
Pre-rinse from set point:
19 l
Rinse steps:
650
Minimum number of steps:
10
Minimum stepping frequency: 50