Measuring system MAK TIGER 3003
45
Measuring System MAK TIGER 3003 Service manual, SA 131128, 10.03.2020
5.18
Mounting of rotation sensor to
rotary pump
Install the proximity switch before the pump inlet is closed as afterwards
the pump can only be turned by hand.
Turn the knob of the rotary encoder disk in front of the desired mounting
opening and screw in the proximity switch to the stop.
Then turn it back by approx. 0.75 to maximally 1 rotation and lock it with
both nuts.
If there is no signal, screw the proximity switch in again with a quarter turn.
Attention:
Non-observance of these mounting instructions can result in the destruction of the
proximity switch.
Cap (possible mount-
ing places)
Disk of rotary encoder
Bottom opening must be open
Locking nuts
Proximity switch
Mount preferably in the
direction of the rear wall