Appendix A: Guidance for Connection to Mains Supply
1. Ensure that the mains feed is from a clean AC supply with the following characteristics:-
Voltage Range
100V to 240V AC (Selectable on Charger)
Frequency
50/60Hz
It is highly recommended that this mains feed is via an appropriate Residual Current Circuit
Breaker, rated at 30mA trip current. This must comply with local regulations.
2. The mains input connectors are as follows:
Device
Input Connector
Mating Cable Connector
8-Way Multiplexing Box
IEC C14
IEC C13
Battery Box
IEC C8
IEC C7
These cables are available in specific versions for most countries.
3. If an un-terminated IEC connector lead is supplied with the equipment, the wires must be
wired accordingly:-
BROWN
LIVE
BLUE
NEUTRAL
GREEN/YELLOW
EARTH
4. It is recommended that the mains to each unit is supplied via an 13A mains fuse:
This is appropriate for the use over the whole supply voltage range.