Model FG1235 Scale Kit
Rev 021210
SECTION I
MODEL FG1235
SYSTEM DESCRIPTION and SPECIFICATIONS
The low power microprocessor circuitry allows the
FG1235 to derive it’s power from the system's DC
battery supply. This eliminates the need for an
external battery charger or the danger of an A.C.
power supply.
SYSTEM DESCRIPTION -
The
FG1235 is specifically designed as a precision scale
system retrofit for the Barton Medical line bariatric
beds. The FG1235 Scale System employs the latest
in microprocessor and load cell technology to
provide accurate and repeatable weight data. Four
(4) identically matched transducers are strategically
placed to ensure an accurate representation of the
patient's weight regardless of weight distribution.
The Patient's weight is displayed on a 16 character
dot matrix liquid crystal display with L.E.D.
backlighting. Patient weight may be viewed in both
POUNDS and KILOGRAMS with the push of a
button with a displayed resolution of 0.1 #/Kg (100
grams). The FG1235 also features a WEIGHT
CHANGE mode which makes it convenient to
monitor the patient's weight relative to ZERO in
order to observe variations in weight. A FREEZE
feature is provided to give the added convenience of
being able to adjust the patient's comfort by adding
or removing a pillow, blanket, or adjusting the chair
without effecting the patient's displayed weight.
1.0
The patient’s weight is displayed on a 16 character
dot matrix L.C.D. With the push of a button, weight
data may be viewed in either POUNDS or
KILOGRAMS with a displayed resolution of 0.1 Lb
or .1Kg.
1.1 INTENDED
USE
The FG1235 is designed for use
as a patient weighing system integrated in the
Barton Medical bariatric bed model _________.
1.2 SPECIFICATIONS
1.2.01
Maximum Weight Capacity -
The FG1235 has a
maximum weight capacity of 1500Lbs. / 680Kgs.
DON NOT exceed the rated weight capacity of the
bed.
1.2.02 Display
Type
- The display is a large 16 Character
Dot Matrix L.C.D. with green L.E.D. back lighting.
Highly visible in all light conditions,
1.2.03 Display
Resolution
– The FG1235 will display
weight data in both POUNDS and KILOGRAMS
with a resolution of 0.1 Lb. or 0.1 Kg.
1.2.04 Accuracy
- The accuracy of the displayed weight is
0.2% +/- 1 digit of the displayed resolution for
calibrated range (Refer to the Calibration Certificate
supplied with the scale for the calibrated range).
1.2.05 Zero
- One button Auto-Zero
1.2.06
Auto Power Down (APD)
– The system will
automatically power down in 60 seconds to
conserve battery life.
1.2.07 Averaging
– An Automatic Digital Filter samples
the weight data to filter motion.
1.2.08 Power
Supply
- 24v DC.
1.2.09
Warranty 1 YEAR LIMITED WARRANTY
-
Each
S
scale system is manufactured
with high quality components. SR Instruments, Inc.
warrants that all new equipment will be free from
defects in material or workmanship, under normal
use and service, for a period of one year from the
date of purchase by the original purchaser. Normal
wear and tear, injury by natural forces, user neglect
and purposeful destruction are not covered by this
warranty. Warranty service must be performed by
the factory or an authorized repair station. Service
provided on equipment returned to the factory or
authorized repair station includes labor to replace
defective parts. Goods returned must be shipped
with transportation and/or broker charges prepaid.
SR Instruments obligation is limited to replacement
of parts which have been so returned and are
disclosed to SR Instruments satisfaction to be
defective. The provisions of this warranty clause are
in lieu of all other warranties, expressed or implied,
and of all other obligations or liabilities on SR
Instruments part and it neither assumes or
authorizes any other person to assume for SR
Instruments any other liabilities in connection with
the sale of said articles. In no event shall SR
Instruments be liable for any subsequent or special
damages. Any misuse, improper installation or
tampering, shall void this warranty.
1.2.10 Calibration
-
Calibration is traceable to NIST
Standards.
1.2.11 Operating
Conditions
– Care has been taken in the
design of this product to provide accurate and
reliable information under normal operating
conditions. When proper care is exercised, your
scale will provide many years of reliable service.
Normal conditions for this product would be :
Temperature Range: 20
o
C
to
30
o
C (68
o
to
85
o
f)
Humidity Range: 0%-95% (non-condensing).
Avoid exposure to high pressure water or steam.
1.2.12
Transport and Storage
– To maintain proper
operation of this instrumentation transport and
storage conditions should not vary outside the
following conditions:
Humidity:
0% to 99%
(non-
condensing).
,
Ambient Temp:
-20
0
C to +50
0
C
1.2.13
Patient Applied Part Classification
– TYPE B
Page 1