7-14
920512-768
Operator Manual
Troubleshooting
15.
If ph Neutralizing System option is
present
, alarm sounds and screen dis-
plays:
ALARM: NEUTRAL. PUMP
MOTOR OVERCURRENT
Printer records:
*ALARM NEUTRAL. PUMP
MOTOR OVERCURRENT
HH:MM:SS YY/MM/DD
Control detects a malfunction with neutralizing pump.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
Neutralizing pump malfunction – Call STERIS.*
16. Alarm sounds and screen displays:
ALARM: EXHAUST FAN
MOTOR OVERCURRENT
Printer records:
* ALARM
EXHAUST FAN
MOTOR OVERCURRENT
HH:MM:SS
YY/MM/DD
Control detects a malfunction with exhaust fan.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
Exhaust fan malfunction – Call STERIS.*
17.
If Drying System option is present
,
alarm sounds and screen displays:
ALARM: DRYING FAN
MOTOR OVERCURRENT
Printer records:
* ALARM
DRYING FAN
MOTOR OVERCURRENT
HH:MM:SS
YY/MM/DD
Control detects a malfunction with drying fan.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
Drying fan malfunction – Call STERIS.*
18. Alarm sounds and screen displays:
ALARM: VHP FAN
MOTOR OVERCURRENT
Printer records:
* ALARM
VHP FAN
MOTOR OVERCURRENT
HH:MM:SS
YY/MM/DD
Control detects a malfunction with VHP fan.
1.
Press
Silence Buzzer
or
ACK
touch pad to silence
alarm buzzer.
2.
VHP fan malfunction – Call STERIS.*
* Service charges may be incurred. Consult your warranty for details.
Table 7-4. Troubleshooting Chart - Alarm and/or Printout (Cont’d)
ALARM MESSAGE
POSSIBLE CAUSE AND CORRECTION