SHANGHAI BAUDCOM COMMUNICATION DEVICE CO.,LTD
- 7 -
5.2
Calibration
Choose “Calibration” tab, the user can perform user self-calibration operation also can
choose “Factory Reset” mode. For example, the current optical power at 1310nm is -10dBm,
type -10dBm in the text box of input standard power, and click Calibration tab, then finished
the calibration operation at 1310nm wavelength. See figure 5.3
5.3
Maintenance
1 .The tester should work without obvious vibration.
2. Always keep clean fiber-optic connectors before using, keep the unit free from oil and
dust.If there is dirt, unscrew the adapter and use wipes or paper and anhydrous alcohol to
clean the end face
。