Maintenance and care
Syscompact 2000 M pro
82 / 98
822-175-2
15.4 Testing the discharge unit
Check the discharge unit works correctly every 12 months or as required. The discharging
resistance of the discharge unit must be 16.5 kOhm (± 10%).
Prerequisites
The resistance measurement is performed before the system is put into operation.
This is because the residual charge in the surge capacitors and the thermal heating of the
resistances influence the resistance measurement.
The system is switched off and is disconnected from the mains voltage.
The system is earthed correctly.
Procedure
1. Connect the ohmmeter to the HV connection cable and the screen of the HV connection
cable.
2. Measure the resistance between the HV connection cable and the screen.
The resistance must be 16.5 kOhm (± 10%).
3. If the measured resistance deviates from 16.5 kOhm by more than 10%, contact your
nearest BAUR representative (http://www.baur.eu/baur-worldwide).