Syscompact 2000 M pro
Maintenance and care
822-175-2
83 / 98
15.5 Checking and cleaning the connection cables and
accessories
WARNING
Dangerous voltage in system components and adjoining live plant
parts
Danger to life, risk of injury due to electric shock.
You may touch the live plant parts and connection fittings that were
under voltage only after they have been discharged and earthed.
Cover the live plant parts properly.
NOTICE
Damage to cable due to aggressive cleaning agents
Do not use any abrasive, corrosive cleaning agents or strong solvents.
Ensure material compatibility.
Do not clean the connection cables with acetone or thinner.
Mild cleaning agents or petroleum ether
Lint-free cleaning cloth
Checking and cleaning after each use
1. Each time after using the system, clean the connection cables.
2. Check the connection cables for damage.
Cracks, breaks or other damage in the connection cable can damage the cable.
3. If dirty, clean the connection accessories with a lint-free cloth.
Dirty or corroded contacts can affect the measurement and are often the cause for device
damage.
Regular inspection
Every four weeks, check the condition of all connection cables. To do this, unwind the
connection cables and examine them for cracks, damage and any dirt.
Regularly check all connection accessories for damage and dirt.
Summary of Contents for Sys compact 2000 M pro