Syscompact 2000 M pro
Maintenance and care
822-175-2
85 / 98
15.7 Replacing the rechargeable batteries of the IRG 2000
NOTICE
Material damage may be caused by having the wrong polarity of the
rechargeable battery
Having the wrong polarity of the battery can damage the electronics.
When inserting the rechargeable battery, pay attention to the polarity of the
battery.
Prerequisites
Phillips screwdriver (size 2)
Rechargeable batteries:
7 x NiMH 2100 mAh Panasonic HHR 210 A
Procedure
1. If the IRG 2000 is used in a system, remove the IRG 2000 from the system.
2. Undo the 6 screws of the battery compartment on the rear side of the device.
3. Remove the cover of the battery compartment.
4. Remove the old batteries.
5. Insert the new batteries.
6. Secure the cover of the battery compartment with the 6 crews.
7. Dispose of the old batteries in compliance with the local regulations.