46
TECHNICAL,
SER
VICE & CARE
47
TECHNICAL,
SER
VICE & CARE
NOTE: If you put the new batteries in the meter
within 5 minutes of taking the old ones out, all your
settings and results will be saved. If you take more
than 5 minutes, you will need to reset the date and
time although your other settings and results will
be saved.
WARNING
Keep batteries away from children. Lithium
batteries are poisonous. If swallowed,
immediately contact your doctor or poison
control center. Discard batteries according to
your local environmental regulations.
Changing the Batteries
Changing the Batteries
1.
Turn off your meter before
changing your batteries.
2.
Press firmly on the battery
cover and slide in the
direction of the arrow.
3.
Remove the old batteries
by pressing the top of the
battery which will raise the
bottom of the battery up.
4.
Place new batteries under
the prongs and into the
battery compartment
with the “+” side up. [Use
two 3-volt (DL2032 or
CR2032) lithium batteries.]
5.
Slide the battery cover
back into place, lining up
with the open slots, and
close firmly.
Changing the Batteries
Batteries should be replaced
when the meter continually
displays
or the meter
briefly shows
and then
turns off
(see Error Codes
and Symbols, page 42).
84721973_CntrPLUS_UG_EN_FpBp_v2.indd 46-47
84721973_CntrPLUS_UG_EN_FpBp_v2.indd 46-47
1/22/13 9:41 AM