Chapter 4: Operating Instructions
4-4
Clinician’s Operation Manual
CAUTION! Due to limitations of the Li-Ion chemistry, the rechargeable battery pack
should not be charged at ambient temperatures above 45 °C (113 °F) or
below 5 °C (41 °F)
CAUTION! The WW1090 rechargeable battery pack is shipped with only 30% of full charge.
The battery pack must be charged completely before use.
CAUTION! The WW1090 rechargeable battery pack utilizes Li-Ion secondary cells. Dispose
of spent batteries in compliance with your institution’s guidelines and local
ordinances.
NOTE: The WW1090 rechargeable battery pack utilizes circuitry that optimizes the
charging of the batteries. New packs will require multiple charge / discharge
learning cycles before optimum performance is obtained.
NOTE: The rechargeable battery can be charged in the oximeter or in the spare bay of the
WW1025 Docking Station. The WW1095 AC power supply is required for charging,
and can be used with both the oximeter and the Docking Station.
External Power
WARNING! Disconnect the external power supply from the monitor before disinfecting or
cleaning the monitor.
WARNING! Do not plug the monitor into an outlet controlled by a wall switch.
WARNING! Do not allow any moisture to contact the AC power supply connectors, or a
safety hazard may result. Ensure that hands are thoroughly dry before handling
the AC power supply.
WARNING! Do not place the monitor in the patient’s bed or crib. Do not place the monitor
on the floor.
WARNING! Failure to place the monitor away from the patient may allow the patient to
turn off, reset, or damage the monitor, possibly resulting in the patient not
being monitored. Make sure the patient cannot reach the monitor from their
bed or crib.
WARNING! Failure to carefully route the cable from the sensor to the monitor may allow
the patient to become entangled in the cable, possibly resulting in patient
strangulation. Route the cable in a way that will prevent the patient from
becoming entangled in the cable. If necessary, use tape to secure the cable.
WARNING! If there is a risk of the AC power supply becoming disconnected from the
monitor during use, secure the cord to the monitor several inches from the
connection.
WARNING! Patient safety can be compromised by the use of a power supply not supplied
by Smiths Medical PM, Inc. Use only the power supply included with your
monitor, or one approved by Smiths Medical PM, Inc.
WARNING! Ensure the device’s AC rating is correct for the AC voltage at your installation
site before using this monitor. The monitor’s AC rating is shown on the external
power supply. If the rating is not correct, do not use the monitor. Contact
the Smiths Medical PM, Inc. service department, or your authorized service
representative, for help.