Chapter 4: Operating Instructions
4-8
Clinician’s Operation Manual
Checking the Monitor’s Configuration
While turning the monitor on, hold the
x
key until
[lr
starts to flash, then release the key
while
[lr
is flashing to see a list of configuration settings.
NOTE! If the
x
key is held for about 15 seconds or more,
[lr
will stop flashing and
TREND DATA WILL BE CLEARED!
Configuration settings are:
Auto Shutoff (
•
as
), default = no;
LED Brightness (
•
db
), default = 8 (of 10);
Trend Interval (
•
se[
), default = 4 seconds;
Oximeter High Sensitivity (
•
ks
) - WW1020 only, default = off
Oximeter Averaging (
•
aug
) - WW1020 only, default = 8 seconds pulse rate, 8 beat SpO
2
;
Language - (
•
lan
) - User selectable
While the language setting is displayed, successive
x
key presses will cycle through the language
selections:
English (
•
En
) - default
French / Français (
•
fr
)
German / Deutsch (
•
de
)
Spanish / Español (
•
es
)
Italian / Italiano (
•
it
)
Continental Portuguese / Português (
•
pt
)
Brazilian Portuguese / Português - Brasil (
•
br
)
Swedish / Svenska (
•
sd
)
Dutch / Nederlands (
•
nl
)
Stop pressing the key when the desired language is selected. After five (5) seconds, normal
monitoring mode resumes.
Checking the Monitor’s Performance
Pulse oximeters do not require user calibration. If checking the function of the device is desired,
an Oximetry Patient Simulator (Smiths Medical PM, Inc. catalog number 1606) is available as an
accessory. The simulator attaches to the oximeter in place of the sensor. It provides a known SpO
2
and pulse rate signal to the oximeter.
NOTE: A Patient Simulator does not calibrate the monitor. The monitor does not require
calibration. A Patient Simulator provides a known SpO
2
value and pulse rate to the
monitor that allows the monitor’s performance to be checked.
NOTE: A Patient Simulator cannot be used to assess the accuracy of a pulse oximeter and/
or sensor.
NOTE:
f
Follow the instructions included with the Patient Simulator.
Summary of Contents for 10
Page 2: ......
Page 6: ...Table of Contents iv Clinician s Operation Manual This page is intentionally left blank ...
Page 81: ......