Chapter 7: PC Communication Setup
Clinician’s Operation Manual
7-1
Chapter 7: PC Communication Setup
Description
Real Time and Trend Data may be sent to a PC from the oximeter in the following ways:
The Data/Power Connector of the oximeter or of the Docking Station may be connected to
•
a PC using the USB Interface Cable (
<
WW1089). Trend data or Real Time patient data will
be downloaded using a proprietary data protocol.
The oximeter sensor can be replaced with the WW3350 or 3339 cable to transfer comma-
•
separated value (.CSV) ASCII Trend Data via RS232.
WARNING! When connecting this monitor to any instrument, verify proper operation
before clinical use. Refer to the instrument’s user manual for full instructions.
Accessory equipment connected to the monitor’s data interface must be
certified according to the respective IEC standards, i.e., IEC 60950 for data
processing equipment or IEC 60601-1 for electromedical equipment. All
combinations of equipment must be in compliance with IEC 60601-1-1 systems
requirements. Anyone connecting additional equipment to the signal input
port or the signal output port configures a medical system, and, therefore,
is responsible that the system complies with the requirements of the system
standard IEC 60601-1-1.
WARNING! IEC 60950 approved equipment must be placed outside the “patient
environment”. The patient environment is defined as an area 1.5 m (4.92 feet)
from the patient.
PATIENT ENVIRONMENT
R 1.5 m (4.9 f
t.)