Chapter 7: PC Communication Setup
7-4
Clinician’s Operation Manual
Sensor / RS232 Port
Trend data may be downloaded in a comma-separated value (.CSV) format to a computer’s serial
(RS232) port.
The following items will be needed:
Oximeter
•
HyperTerminal or other PC communication software
•
WW3350 Printer Adaptor cable
•
3339 PC Adaptor cable
•
Figure 7-3: Connecting to the RS232 Port
How to Set Up Equipment
Refer to Figure 7-3 for the cable connections.
Setup the PC communication program to accept the following RS232 data format:
1.
Data Type: ASCII
•
Data Format: 9600 baud, 1 start bit, 8 data bits 1 stop bit, no parity
•
Select any available RS232 Com Port (typically com1 or com2)
•
Turn off the oximeter.
2.
Remove the oximetry sensor from the sensor connector.
3.
Connect the printer adaptor cable (WW3350) to the sensor connector. Make sure the “BCI”
4.
end is connected to the oximeter.
Connect the PC adaptor cable’s DB-9 connector to the mating connector of the printer cable
5.
labeled “Printer”.
Connect the other end of the PC adaptor cable to the PC’s RS232 Com Port selected in step 1.
6.
Turn on the oximeter to start sending all of the stored trend data to the PC.
7.
NOTE!
1
Review the PC communication program’s instructions to determine how to
save the data to the PC.
Connect to Oximeter
Cable (
<
WW3350)
Printer Adapter
DB-9 Connector
PC Adapter
DB-9 Connector
Cable (
<
3339)
Connect to IBM
Compatible PC
or Printer