Chapter 3: Block Level Description
Advisor
®
Service Manual
3-3
non-Invasive Blood Pressure (nIBP) Assembly
The Non-Invasive Blood Pressure (NIBP) Assembly consists of three components, each mounted separately
below the Main Board: the pump, the valve / manifold assembly, and the NIBP Board assembly. The NIBP Board
assembly connects to the Main Board via connector (J2), sup12VDC power and bi-directional TTL serial
communications directly to the ARM7 microprocessor. Inverter IC (U12 on board 80002B1) provides 3.3 to 5.0 V
logic translation for the +5V logic needed by the NIBP Board.
LCD Backlight Inverter Board
WARnInG! The LCD Backlight Inverter Board operates at voltages in excess of 1000 VAC and may arc to
items in close proximity.
The LCD display panel is backlit with cold cathode filament lamps powered by the LCD Backlight Inverter
Board. The board co12VDC to an approximate 44kHz, 1400 VAC drive for the lamps. The LCD Backlight
Inverter Board is mounted directly below the LCD panel, sandwiched between the front half of the case and the
internal mounting plate for the LCD and Main Board. The Inverter Board is connected to the Main Board by way
of connector J6. The ARM7 microprocessor has direct control of switching power to the LCD Backlight Inverter
Board by way of mosfets Q6 & Q7.
Printer Adapter Board (Printer option)
When the optional graphics printer is installed, a small Printer Adapter Board connects directly to the printer and
is secured to the same internal mounting bracket as the printer. The Printer Adapter Board also provides a ribbon
cable connector for connection to the Main Board as well as a large filter capacitor to help supply peak currents
required by the thermal print-head when printing. The Printer Adapter Board assembly connects to the Main
Board via connector (J21), sup12VDC power, +5VDC power, and bi-directional TTL serial communications
to the ARM7 microprocessor via UART (U74).
Co
2
Module (optional)
The CO
2
Module can be installed on the back of the monitor. It is based on Smiths Medical PM 1410 Gas Module
(see Smiths Medical PM Document 57700TEC for description of 1410 Module).
Module attachment to the monitor in the advanced mode is shown on the drawing (see 57692A2).
The round mini-DIN connector located on the back plate of the monitor supplies power and serial
communication lines to the Module. Inside the monitor, mini-DIN connector is connected to J4 (power) and J6
(communications) on the Main Board.
As long as 1410 Module is installed, Main Monitor Software recognizes it and enables CO
2
parameter box and
CO
2
waveform.
Summary of Contents for Advisor
Page 2: ......
Page 6: ...Table of Contents iv Advisor Service Manual This page is intentionally left blank ...
Page 8: ...Revision History vi Advisor Service Manual This page is intentionally left blank ...
Page 43: ...Chapter 4 Diagrams Advisor Service Manual 4 1 Chapter 4 Diagrams ...
Page 44: ...Chapter 4 Diagrams 4 2 Advisor Service Manual ...
Page 45: ...Chapter 4 Diagrams Advisor Service Manual 4 3 ...
Page 46: ...Chapter 4 Diagrams 4 4 Advisor Service Manual ...
Page 47: ...Chapter 4 Diagrams Advisor Service Manual 4 5 ...
Page 48: ...Chapter 4 Diagrams 4 6 Advisor Service Manual ...
Page 49: ...Chapter 4 Diagrams Advisor Service Manual 4 7 ...
Page 50: ...Chapter 4 Diagrams 4 8 Advisor Service Manual ...
Page 56: ...Appendix App 2 Advisor Service Manual This page is intentionally left blank ...
Page 57: ......