24
HTS Customer Care Kits User’s Guide
NOTE
Use caution when handling the sample injection port assembly. This
assembly is fragile and can be easily damaged.
Replacing Tubing Assembly Components
1
Turn off the cytometer and HTS and move the sample probe into the
sample injection port.
2
Remove the sample probe tubing assembly.
3
Unscrew the sample injection tubing from the probe and the top of primary
pump.
The sample injection tubing connectors are indicated in the following
figure.
Clamp
Primary pump
Tubing assembly