background image

Alaris™ VP Plus Guardrails™ Volumetric Pump
Model: 9003TIG03-G

Quick Reference Guide – en

0000CF04424.indd   1

17/09/2019   10:44:59

Summary of Contents for 9003TIG03-G

Page 1: ...Alaris VP Plus Guardrails Volumetric Pump Model 9003TIG03 G Quick Reference Guide en 0000CF04424 indd 1 17 09 2019 10 44 59...

Page 2: ...DFUs are the sole reference for Pump information This guide is intended only for qualified healthcare professionals having received appropriate training to operate the Pump the HCPs HCPs must have con...

Page 3: ...ard i PRIME BOLUS button Press to access PRIME or BOLUS softkey Press and hold down softkey to operate d OPTION button Press to access optional features e PRESSURE button Use this button to display th...

Page 4: ...or is difficult to close open the door fully to verify correct loading of the infusion set 1 Close roller clamp 2 Insert the bag spike into the fluid container and hang appropriately at a minimum heig...

Page 5: ...7 Close door and open roller clamp Ensure no drops are falling in the drip chamber If drops are falling in the drip chamber then clamp the infusion set immediately open door and verify correct loading...

Page 6: ...OFILE ICU 3 When PROFILE screen is shown Select NO to change profile and then follow on screen prompts Select YES retains current profile Note The PROFILE screen is only shown if more than one profile...

Page 7: ...ess OK to confirm Set the END RATE using the f keys Press OK to confirm VOLUME RATE 0 0ml h ON HOLD PRIMARY SET RATE WITH VTBI 8 Enter or adjust the rate using the f keys if required 9 Connect the ful...

Page 8: ...ands Free Bolus 1 During infusion press the i button once to display the BOLUS MODE screen 2 Press the YES softkey to go to the HANDSFREE bolus screen or press HANDSON softkey to go to the HANDSON onl...

Page 9: ...larm limit set through the Auto pressure or manual functions can assist in optimizing the time to alarm Despite the benefits provided by early detection of occlusions and reduced time to alarm the Pum...

Page 10: ...cally on the display screen Auto Pressure The Auto Pressure feature may be used when a stable pressure has been achieved over a short period of infusion If Auto Pressure has been enabled in the editor...

Page 11: ...e up to the maximum occlusion alarm limit defined within the profile Pressure Trend TREND 71mmHg BACK 17 15 0 1 To check the pressure trend graph press the e button 2 Press the TREND softkey to view t...

Page 12: ...icy Check fluid level in container Check fluid level in drip chamber Restart infusion DOOR OPEN Door was opened during an infusion Close door or clamp infusion set using roller clamp Restart infusion...

Page 13: ...and ensure the infusion set is loaded correctly Consider if the 15 m filter in the drip chamber is blocked or partially blocked Ensure the infusion set has not been used for more than 72 hours If alar...

Page 14: ...ce flow sensor or set VTBI WRONG SET Safety clamp not detected Clamp infusion set using roller clamp Check set and close door Replace infusion set as required DOOR CLOSE INCOMPLETE Safety clamp in non...

Page 15: ...t Check fluid level in drip chamber Check fluid level in container SET CLOCK Date time not set Qualified service personnel must set date time Press CANCEL softkey to continue TITRATION Rate titration...

Page 16: ...ET VTBI No flow sensor VTBI Set VTBI or fit flow sensor SET NOT FITTED No infusion set fitted Fit infusion set LOCKED Rate change attempted whilst locked Unlock rate to adjust infusion settings ADD DR...

Page 17: ...Concentration entered exceeds the hard maximum limit or is under the hard minimum limit BOLUS DOSE NOT PERMITTED Bolus dose rate entered exceeds the hard maximum limit WEIGHT ABOVE LIMIT Patient weig...

Page 18: ...ails are trademarks of Becton Dickinson and Company or its affilates 2019 BD All rights reserved BD Switzerland S rl Route de Crassier 17 Business Park Terre Bonne Batiment A4 1262 Eysins Switzerland...

Reviews: