BDDF00160 Issue 1
33/40
Alaris™ GW Volumetric Pumps
Specifications
Specifications
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24 .
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24 .
Electrical Safety
IEC/EN 60601-1 - Typical earth leakage current 40µA .
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation .
Dielectric Strength
Proof strength test 1 .7kV dc (live and neutral to earth) for 10s
Performance strength test 500V dc (live and neutral to earth)
Proof Strength Test
The proof strength test is applied at the factory . It is not recommended that the proof strength test is reapplied if the Pump is tested
again during service .
Classification
Class I Equipment . Continuous, Portable Equipment, type 4 .
AC Power Supply
220 - 240 VAC, 50 - 60Hz, 10VA (nominal) .
Protection against fluid ingress
IP32 - Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater than 2 .5mm .
Dimensions
137mm (w) x 140mm (h) x 105mm (d) . Weight: approx . 1 .5kg (excluding power cable) .
Environmental Specifications
Condition
Operating
Transport and Storage
Temperature
+5°C - +40°C
-20°C - +50°C
Humidity
20% - 90%*
10% - 100%*
Atmospheric Pressure
700hPa - 1060 hPa
500hPa - 1060hPa
Priming the Infusion set / Prime
Parameter
Range
Priming Rate
Fixed:>999ml/h
Priming Volume
0 - 40ml***
Starting the Infusion / Set-up
Infusion Parameter
Micro
Standard
Flow Rate
1 .0 - 99 .9ml/h**
1 - 999ml/h***
VTBI
0 .1 - 99 .9ml**
100 - 999ml***
1 - 9999ml***
VI
0 .0 - 99 .9ml**
100 - 9999ml***
0 - 9999ml***
Administering a Bolus
Parameter
Range
Bolus Rate
1 - 999ml/h***
Bolus Volume
0 - 99ml***
Max Bolus Volume after release of hard occlusion
<0 .6ml
*Non condensing .
**Measured in 0 .1ml increments .