0000CF04392 Iss. 1   7/12

User Selectable Options

The pump must be on HOLD. Press 

d

.

Check the Battery Status

Note:

 

Only available when Pump is operating on battery power.

Press and hold 

e

 for 2 seconds until Battery Status is displayed:

,

 

• Battery has greater than 30 minutes run time.

.

• Battery is low and has approximately 30 minutes or less 

run time.

Press 

e

 button to go to next option.

Selecting the Occlusion Pressure Level (if Enabled)

PrES

 option is displayed:

s

• Change occlusion pressure level to high (

HI

), normal (

nor

or low (

lo

) using 

h

 / 

f

.

• Press 

e

 to confirm and go to next option.

Selecting the Alarm Volume (if Enabled)

tonE

 option is displayed:

t

• Change alarm volume between 1 and 7* using 

h

 / 

f

.

• Press 

e

 to confirm and go to next option.

* 1=Low, 7=High

Selecting VTBI / Time Infusion (if Enabled)

CLoc

 option is displayed:

q

• Select 

On

 or 

OFF

 using 

h

 / 

f

 keys to turn the VTBI/time 

infusion** setting on or off.

• Press 

e

 to confirm and go to next option.

** Ensure 

TIME

 indicator is lit, if set to 

On

.

Selecting Micro Mode (if Enabled)

0.0

 option is displayed:

r

• Select 

On

 or 

OFF

 using 

h

 / 

f

 keys to turn micro mode*** 

on or off.

•  Press 

e

 to confirm.

*** Ensure 

MICRO

 indicator is lit, if set to 

On

.

Configur

able Set-up Mode

Summary of Contents for Alaris 800TIG2GBD1

Page 1: ...Alaris GW 800 Volumetric Pump Models 800TIG2GBD1 800TIG2RWN1 Quick Reference Guide en ...

Page 2: ...plete pump information refer to the Alaris GW 800 Volumetric Pump Directions For Use All persons operating the pump should consult the Directions For Use PRESSURE SENSOR TUBING guide DOOR latch AIR SENSOR FLOW STOP DISPLAY INDICATORS MAIN DISPLAY ENTER CHEVRONS RUN HOLD ON OFF PRIMARY SECONDARY PRIME BOLUS CLEAR MUTE MUTE INDICATOR ALARM INDICATOR Controls ...

Page 3: ...te To re enable the alarm audio press any other button or key to unmute PRIME BOLUS button Primes the Infusion set Administers bolus during the infusion PRIMARY SECONDARY button Switches the Pump between Primary and Secondary infusion modes if enabled ENTER button Scrolls between rate time VTBI and total volume infused VI Enters values for selected infusion configuration parameters Confirms the ra...

Page 4: ...the Volume To Be Infused VTBI in millilitres ml VI When illuminated the Pump is displaying the Volume Infused VI in millilitres ml TIME When illuminated the Pump is displaying the infusion time in hours mins MICRO When illuminated the Pump is operating in the Micro Mode When not illuminated the Pump is in the Standard Mode SEC When illuminated the Pump is operating in the Secondary Mode When not i...

Page 5: ...oor 3 Release flow stop mechanism by pushing the lever arm up and to the right 4 Insert the infusion set from left to right into the pumping channel Avoid any slack and avoid stretching or pulling the tubing 5 Engage the flow stop mechanism by pushing the lever left and down 6 Close door Press recesses firmly as shown to ensure latch clicks into place Loading the Infusion Set ...

Page 6: ... visible air is cleared from the line 3 Connect the infusion set to the patient access device Enter rate using h f Press e once to confirm rate and scroll to VTBI 4 Enter VTBI using h f or if using a flow sensor VTBI can be switched off by scrolling until OFF is displayed Press e once to confirm VTBI and scroll to VI 5 If necessary press c to clear VI 6 Press d to start the infusion Starting the P...

Page 7: ... to high HI normal nor or low lo using h f Press e to confirm and go to next option Selecting the Alarm Volume if Enabled tonE option is displayed t Change alarm volume between 1 and 7 using h f Press e to confirm and go to next option 1 Low 7 High Selecting VTBI Time Infusion if Enabled CLoc option is displayed q Select On or OFF using h f keys to turn the VTBI time infusion setting on or off Pre...

Page 8: ...ration To change the rate while infusing if titration enabled 1 Press h f to adjust rate 2 Press e to confirm change of rate Administer BOLUS Press b To administer a bolus while infusing Press b While boL is displayed press b again and hold to administer the desired bolus volume Features ...

Page 9: ... L Door open Close door and restart infusion M System fault Switch pump off Remove pump from service and have the pump inspected by a qualified service engineer N Flow error Flow sensor attached to secondary infusion set Close in line clamp Ensure tubing is loaded correctly Ensure ample fluid is in container Clear blockage Close pump door Restart infusion Ensure flow sensor is attached to primary ...

Page 10: ...ed pre set volume to be infused with KVO switched off Set new VTBI F Low battery At least 30 mins before K alarm Connect pump to AC power source H The pump has been left unattended for two minutes and infusion has not started Attend to pump Alarms continued ...

Page 11: ...key to continue the infusion once prescribed bolus has been administered D VTBI limit reached Pump will infuse at the KVO rate until d is pressed E The pump is priming the infusion set Ensure all air has been primed out of the infusion set before starting the infusion G The pump is on hold Press d to restart continue infusion or press e to return to setup I Automatic infusion set check Allow test ...

Page 12: ... BD Logo and Alaris are trademarks of Becton Dickinson and Company or its affiliates 2019 BD All rights reserved BD Switzerland Sàrl Route de Crassier 17 Business Park Terre Bonne Batiment A4 1262 Eysins Switzerland ...

Reviews: