Loading the syringe:

1.  Pull syringe barrel clamp out, rotate 

it to the left and gently release it.

2.  Twist gripper control clockwise to raise  

the drive head 

(gray)

 to its fully extended position.

3.  Insert the syringe by sliding the flat edges between  

the barrel flange grippers 

(see drawing)

.

4.  Lock syringe in place by closing the barrel clamp.  

To close the  barrel clamp, pull out and rotate it to right. 

5.  Twist gripper control clockwise, gently lower drive head  

and release gripper to lock the plunger in place.

Priming the syringe using the Alaris

TM

 PCA module:

Note: Do not prime the syringe while it is attached to  
the patient.

1.   

PRIME

 softkey is available only after syringe type and 

medication selection 

(prior to infusion mode selection).

2.   At the Infusion Mode screen, press 

OPTIONS,

 then press  

PRIME SET WITH SYRINGE.

3.   Press and hold 

PRIME

 key to prime tubing.

Programming guide

 

initial setup:

1.   Load the syringe with the administration set attached. 
2.   Press the 

SYSTEM ON

 key.

3.   Select 

Yes

 or 

No

 to 

New Patient?

4.  Select appropriate profile.
5.   If required, enter patient identifier or press 

EXIT

.

6.  Press the 

CHANNEL SELECT

 key.

7.  Set key to 

PROGRAM

 position.

8.  Confirm time of day or change time if necessary.
9.  Select the correct syringe type and size. Press 

CONFIRM

Note:

 

Applies to all syringes.

 

Programming the Alaris PCA module with PCA Pause  
Protocol enabled:

1.   Perform steps 1–10 of initial setup and then continue with  

the following steps.

2.   Review the Clinical Advisory: 

Attach an Alaris SpO

2

 or Alaris 

EtCO

2

 module now.

3. Press 

CONFIRM.

 

Note:

 

If a monitoring module is not attached and started,  

PCA Pause Protocol will 

not

 activate.

4.  Press the 

NEXT

 key to verify medication parameters.

5.  Review Advisory.
6. Press 

PAUSE LIMITS

 to review the settings or  

press 

CONFIRM.

7.  Select desired infusion mode and follow on-screen 

programming prompts.

Alaris

TM

 PCA module pocket guide 

bd.com

© 2017 BD. BD, the BD Logo and Alaris are trademarks of Becton,  

Dickinson and Company. IF7879 (0917/2500) CF/201/17/0043

4.  Set key to the Program position and close door.
5.  Press the 

CHANNEL SELECT

 key.

6.  Select and confirm correct syringe type and size.
7.  If using the same drug, dosing units and concentration,  

then press 

RESTORE.

8.  Verify that the drug, concentration and dosing parameters  

are correct.

9.  Lock door and open tubing clamp.
0.  Review settings and press 

START.

Change program/mode:

1. Press 

CHANNEL SELECT

 key.

2. Press 

PROGRAM.

3.  Set key to Program position or enter authorization  

code 

(if enabled).

4. Press 

CHANGE MODE.

 Select desired infusion mode  

and follow on-screen prompts.

Give a clinician bolus dose:

1. Press 

CHANNEL SELECT

 key.

2. Press 

BOLUS DOSE.

3.  Set key to Program position or enter authorization  

code 

(if enabled).

4.  Enter the bolus dose amount and lock the door. 
5. Press 

CONFIRM.

6.  Review settings and press 

START.

Beginning of shift/Summary review:

1. Press 

CHANNEL SELECT

 key and verify settings.

2. Press 

START

 key.

Access drug event history:

1. Press 

CHANNEL SELECT

 key.

2. Press 

OPTIONS;

 then press 

DRUG EVENT HISTORY.

3. Press 

EXIT

 and then press 

START.

Warning: Before loading or 
unloading

 the syringe, always turn 

off fluid flow to the patient using a 
tubing clamp. Uncontrolled fluid flow 
can occur when the administration 
set is not clamped or turned off and 
may cause serious injury or death.

1. Press 

PAUSE

 on PCA module. 

Note:

 Always clamp administration 

set before changing syringe.

2.  Use the PCA key to unlock  

the door; remove the syringe  
and press 

SILENCE.

3.  Attach the new syringe to the 

tubing and load the new  
syringe into the PCA module.

Change syringe and use Restore feature

Operator precautions: For proper operation of the Alaris

 

System, the user must be familiar with the features, disposables, 

administration sets, setup and programming.

This guide includes selected information and suggestions and is not intended to be comprehensive instructions for the setup 

and operation of the Alaris System. For complete instructions along with warnings and cautions, refer to the User Manual.

10.  Select correct medication and concentration.
11.  At Infusion Mode screen, press 

OPTIONS

, then press  

  

PRIME SET WITH SYRINGE

.

12.  Press and hold the 

PRIME

 key to prime tubing. 

  

Note:

 Do 

not

 prime the tubing while it is attached to the patient. 

13.  Press 

EXIT

 when priming is complete.

14.  Select desired infusion mode and follow on-screen prompts to  

  enter dosing parameters.

15.  Close and lock the door. 
16.  Attach administration set tubing to the patient.
17.  Review and verify settings.
18.  Open tubing clamp.
19.  Press 

START

.