58
BD CARV II Installation and User’s Guide
NOTICE
Allowable values range from 450 to 1050. Any or all positions of one
device can be calibrated in one command. Calibration is used to align each filter
position to the light path.
Improper calibration will negatively affect the optical
performance of the CARV II.
f
– Read Calibration Values of All Filter Wheels
NOTICE
The calibration values for each filter position of each filter wheel are
reported on separate lines.
q
– Query State of All Devices
Format
f<CR>
Parameters
<device ID>
ID assigned to device (see Table 4-1)
<position>
Position of device that is calibrated to
<value>. For example, A=1, H=8
<value>
Calibrated value for <position>
Return Value
F<device ID>:<position><value><position><value> . . .<CR>
<device ID>:<position><value><position><value> . . .<CR>
<device ID>:<position><value><position><value> . . .<CR>
Examples
f<CR>
J
fA:A450B575C708D825E893F956G1050<CR>
B:A450B582C715D808E895F966G1050<CR>
C:A450B596C702D823E876<CR>
Format
q<CR>
Parameters
<A>
Device ID and position of Excitation Filter Wheel
<B>
Device ID and position of Emission Filter Wheel
<C>
Device ID and position of Dichroic Filter Wheel
<D>
Device ID and position of Confocal Disk Slider
<P>
Device ID and position of Prism Slider