background image

1.  Shut down the BD Accuri C6.
2.  Unscrew the BD Accuri C6 SIP collar, remove the sample stage and install the Low-Profile BD CSampler SIP collar (Figures  

1 - 3). The low-profile BD CSampler SIP collar ensures that collisions do not occur during operation of the BD CSampler.

Figure 1

Figure 2

Figure 3

BD CSampler™ Installation Guide

BD CSampler Accessory Kit

Shipping Contents

• 

BD CSampler Module

• 

BD CSampler Mat

• 

Mounting Bolts (3)

• 

Installation and Removal Tool

• 

BD CSampler SIP Collar

• 

BD CSampler Software CD 

(not shown)

• 

24-tube rack 

(not shown)

Install and validate the BD Accuri™ C6 Flow Cytometer with 6- and 8-peak beads prior to installing the BD 
CSampler.

 

Inspect the BD CSampler Accessory Kit contents to ensure delivery of all components.

BD CSampler Accessory Kit

BD CSampler 

Module

 

Mat

 

Tool

 

BD CSampler 

SIP Collar   

 

Bolts

 

WARNING

 

 

Failure to shut 

down the BD Accuri C6 during BD 

CSampler installation could result in 

damage to either or both electronic 

modules.

Low-Profile BD CSampler

SIP collar

BD C6 SIP collar

 www.AccuriCytometers.com                                 For Research Use Only. Not for use in diagnostic procedures. Class 1 Laser Product                                                  7820020  Rev-3