Overview
The BD
®
Small Particle Detector (SPD) is a side-scatter detector optimized to enhance resolution of small
particles such as extracellular vesicles. This device can be included as an option on a new or existing system.
Note:
The SPD is a separate device installed inside the BD FACSymphony™ A1 flow cytometer, and the output is
captured on the blue laser detector array. For more information, see the blue laser configuration map in
configuration polygon maps (page 101)
. For small particle analysis, low background noise level is critical. The
system must have a low event rate to ensure that the target cells are separated from noise.
SPD components
Item
Description
1
Beam splitter
2
Unredirected laser beam
3
Redirected laser beam
4
Pinhole
5
Bandpass filter
6
Small particle PMT with integrated preamplifier
Beam splitter
The 488 90/10 nm beam splitter directs 90% of the 488 nm scattered light towards the pinhole.