Setting up PMT voltages for small particles
This topic assumes that you have created an SPD QC experiment as detailed in the preceding topic.
Note:
Cytometer settings used for the following procedure are for SPD alignment.
1. If the SPD QC experiment created using the previous procedure has been closed, reopen it.
2. Vortex a Megamix-Plus SSC beads vial and add 500 µL of the bead solution to a clean BD FACS™ tube.
3. Install the Megamix tube and acquire the sample on LOW.
4. Adjust the voltage for BB515 parameter to display bead populations in the middle of the SP SSC-H vs
BB515-H dot plot.
110
BD FACSymphony™ A1 Flow Cytometer User's Guide