4. Acquire and observe the threshold rate for 5 minutes.
a. If the event rate stops decreasing, record 1 minute of data for the filtered diluent on LOW.
b. If the event rate does not decrease, see
Small particle detector troubleshooting (page 116)
.
Aligning the small particle detector
The New Focus Picomotor application can be used to align the SPD if needed.
To use the New Focus Picomotor Application:
1. Open the New Focus Picomotor Application by double-clicking the icon on the desktop.
2. In the Controller area in the upper left of the dialog, select the Controller, as shown in the following screen
capture.
3. In the Motor area, you can select the motors that are configured for your system. Typically, Motor 1 is for
the X direction and Motor 2 is for the Y direction.
4. The Position Status shows the relative position of the motors. Zero the positions so you have a reference
established for the starting point. If you cannot hear Motors 1 and 2 responding, return to the preceding
step and select Motors 3 and 4.
5. Select Relative mode and set the steps value to 1000, as highlighted in the following screen capture.
Chapter 10 Small Particle Detector
113
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...