6. In the Relative steps area, you can select – (minus) or + (plus) to move the motor stage in small increments.
7. While acquiring your MegaMix-Plus bead mixture and examining the SP SSC-H v BB515-H plot and SP-SSC-H
histogram, move the motors one step at a time. The SPD is aligned when the bead peaks are tight and well
separated.
The following plot diagrams and statistics views show an example of a poorly aligned SPD:
114
BD FACSymphony™ A1 Flow Cytometer User's Guide