Status screen
The status screen display is a touchscreen color LCD screen which displays operational data while the system is
running and toggles between two different displays: Normal Mode and Setup Mode.
Normal Mode displays the current settings of the Fine Adjust Potentiometer, the three programmable
potentiometer Set Points, Sheath Tank Pressure Status, Sheath Level, and Waste Level while the system is
running.
The Normal Mode screen is shown in the following illustation.
SETPOINT 1
SETPOINT 2 SETPOINT 3
SETPOINT
FUNCTION
SETUP
MODE
FINE ADJUST
PRESSURE
HIGH
GOOD
LOW
SHEATH
WASTE
6
5
4
3
2
1
7
Number
Definition
1
Fine Adjust.
Shows the current setting of fine adjust. The slider enables quick changes to the potentiometer
setting by moving the slider up or down. Fine adjust can be set in increments of 10 from 0 to 500. The
normal set point is 250. The last value persists, even after cytometer shutdown.
2
Sheath Level.
Shows range from 0 to 100. The display line decreases from top to bottom in sequences of
20% from full level. System status turns yellow at 20%, and red at 0%.
3
Waste Level.
Shows range from 0 (empty) to 100 (full). The display line increases from bottom to top in
sequences of 20%. System status turns yellow at 80%, and red at 100% full.
4
Setpoint.
Stores up to three potentiometer setpoints. A setpoint is set by tapping the Setpoint Function
button, moving the Fine Adjust slider up or down to the desired point, and tapping one of the setpoint
buttons. The Setpoint enables you to store three different settings and quickly change between them.
5
Setpoint Function.
Button enables the setting of a Setpoint.
6
Setup Mode.
Button to enter Setup Mode (Setup Mode can also be entered by pushing the Mode button on
the control panel).
7
Alarm Silenced Indicator.
When the alarm sounds are silenced, the indicator will turn on.
18
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...