Replacing the waste tank cap
Introduction
This topic describes how to replace the waste tank cap. Replace the cap once a month.
Biological precautions
Contact with biological specimens and materials can transmit potentially fatal disease.
To prevent exposure to biohazardous agents:
l
Put the cytometer in standby mode before disconnecting the waste tank to avoid leakage of biohazardous
waste.
l
The waste tank can become pressurized when the cytometer is running. Always disconnect the waste tank
from the cytometer before you empty it. Wait at least 30 seconds for pressure to dissipate before you
remove the waste cap or sensor.
l
Expose waste tank contents to bleach (10% of total volume) for 30 minutes before disposal.
42
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...