Managing background noise
Reducing background noise
The procedure in this topic should be performed at least once after the previous setup procedures, to remove
traces of the MegaMix-Plus beads before checking the background noise. The following procedure assumes that
the system is currently running on LOW with a tube of MegaMix-Plus mixture or filtered diluent on the SIP.
1. Prepare a tube with 3 mL of a 1.5% dilution of BD
®
Detergent Solution Concentrate. See
.
Never mix BD
®
Detergent Solution Concentrate and bleach because they create chlorine gas.
2. Change the tubes on the SIP:
a. While still running the system on LOW, remove the tube containing MegaMix-Plus beads or filtered
diluent.
b. Wipe the SIP with a delicate task wipe.
c. Move the tube support arm to the side for 10 seconds.
d. Install the tube containing detergent solution on the SIP.
To avoid adverse fluctuations in background event rate, ensure that the tube support arm is to
the side for no longer than 15 seconds.
e. Place the tube support arm under the new tube.
3. Continue to run the detergent solution on LOW for 15 minutes. During this time, prepare a clean tube
containing 3 mL of filtered diluent—see
.
4. Change the tubes on the SIP:
a. While still running the system on LOW, remove the tube containing detergent solution.
b. Wipe the SIP with a delicate task wipe.
c. Move the tube support arm to the side for 10 seconds.
d. Place a tube containing 3 mL of filtered diluent on the SIP.
To avoid adverse fluctuations in background event rate, ensure that the tube support arm is to
the side for no longer than 15 seconds.
e. Place the tube support arm under the new tube.
5. Continue to run the filtered diluent for 1 minute on LOW.
Analyzing background noise
To confirm that the background noise has stabilized, the following procedure should be performed after the
previous noise-reduction procedure. In other words, the SPD QC experiment is open, a tube of filtered diluent is
loaded on the SIP, and the system is running on LOW.
1. Move the pointer to a new tube with no data recorded.
2. In the Cytometer window, select the
Threshold
tab.
3. Change the parameter to
SP SSC
and set the threshold value to
200
.
112
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...