Small particle detector troubleshooting
Event rate or background noise is higher than expected
Possible causes
Recommended solutions
Your sample diluent may be
the source of the noise or
there may be leftover beads
in the system.
1. Acquire filtered diluent on LOW and observe event rate for 5 minutes. If event rate is
decreasing, continue to run until desired rate is achieved or rate stops falling.
2. Repeat the noise-reduction procedure—see
Reducing background noise (page 112)
.
3. Prepare a clean new tube of filtered diluent.
4. While still running on LOW, remove the tube of filtered diluent used in step 1.
5. Wipe the SIP with a delicate task wipe.
6. Move the tube support arm to the side for 10 seconds.
7. Place the new tube of filtered diluent on the SIP.
To avoid adverse fluctuations in background event rate, ensure that the
tube support arm is to the side for no longer than 15 seconds.
8. Place the tube support arm under the new tube of filtered diluent and continue to run
on LOW.
STANDBY or PRIME mode is initiated in the middle of the small particle
experiment
Possible causes
Recommended solutions
Tube support arm is left to
the side for longer than 15
seconds
1. Acquire filtered diluent on LOW and observe event rate for at least 2 minutes.
2. Check the background event rate.
3. Continue running filtered diluent until the event rate stops decreasing.
116
BD FACSymphony™ A1 Flow Cytometer User's Guide
Summary of Contents for FACSymphony A1 Flow
Page 1: ...FACSymphony A1 Flow Cytometer User s Guide 23 23437 01 2022 07 For Research Use Only ...
Page 6: ......
Page 10: ...This page intentionally left blank ...
Page 24: ...This page intentionally left blank ...
Page 36: ...This page intentionally left blank ...
Page 50: ...This page intentionally left blank ...
Page 87: ...More information l Running a performance check page 55 Chapter 7 Technical overview 87 ...
Page 88: ...This page intentionally left blank ...
Page 96: ...This page intentionally left blank ...