User’s Guide
70
7. In the
Setup Control
window, select
Check Performance
from
the
Characterize
menu.
8. Click
Run.
9. Ensure that Fine Adjust is set to 250, press
Run
, and
Low
.
Plots appear under the Setup tab and the performance check is
run. The performance check takes approximately 5 minutes to
complete.
10. Once the performance check is complete, click
View Report.
11. Verify that the cytometer performance passed.
In the Setup tab, the cytometer performance results should
have a green checkbox displayed and the word
Passed
next to
it.