User’s Guide
82
Compensation control tubes are added to the experiment.
Worksheets containing appropriate plots and gates are added
for each compensation tube.
Recording
compensation
settings
To record compensation settings:
1. Press RUN and HIGH on the cytometer fluid control panel.
2. Install the unstained control tube onto the SIP.
3. Expand the
Compensation Controls
specimen in the
Browser
.
4. Set the current tube pointer to the unstained control tube (it
becomes green), then click
Acquire Data
in the
Acquisition
Dashboard
.
5. Verify that the population of interest is displayed appropriately
on the FSC vs SSC plot and adjust voltages if necessary.