Chapter 4: Maintenance
49
When to perform
maintenance
procedures
Perform maintenance procedures in the following frequencies.
Cleaning the fluidics
Introduction
This topic describes how to perform the daily fluidics cleaning.
Overview
Cleaning the fluidics daily prevents the sample injection tube from
becoming clogged and removes dyes that can remain in the tubing.
In addition to daily cleaning, follow this procedure immediately
after running viscous samples or nucleic acid dyes such as Hoechst,
DAPI, propidium iodide (PI), acridine orange (AO), or thiazole
orange (TO).
Procedure
To clean the fluidics:
1. Press RUN and HIGH on the cytometer fluid control panel.
2. Install a tube containing 3 mL of a cleaning solution on the SIP
with the support arm to the side (vacuum on) and let it run for
1 minute.
Frequency
Maintenance procedure
Daily
Cleaning the fluidics (page 49)
Shutting down the cytometer (page 51)
Scheduled (every
two weeks)
Periodic (frequency
depends on how
often you run the
cytometer)
Changing the sheath filter (page 56)
Changing the Bal seal (page 58)
Changing the sample tube O-ring (page 60)