Using BD FocalPoint™ Slide Profiler Reports
115
5
The
Review
action indicators.
A black box indicates that the corresponding slide has the action
Review/Process Review/
Rerun/Empty
. A
review
slide must be reviewed on the BD FocalPoint™ GS Review Station.
Process Review
and
Rerun
slides must be manually screened.
6
The results for each slide. This includes:
•
Action required—
Review/Process Review/Rerun/Empty
. (For additional information, see
•
Specimen Adequacy
•
Squamous cells—Detected/Insufficient
•
Endocervical cells—Detected/Insufficient
•
Inflammation/Obscuration—less than 50% (Satisfactory), between 50–75% (SBLB), or
greater than 75% (Unsatisfactory).
•
Ranking information. All successfully processed slides having an action of
Review
or
QC
Review
receive a group (or quintile) ranking, an individual ranking within the print set,
and each individual rank expressed as a percentage of the total of all
Review
slides.
•
For the
group
ranking, slides are assigned to a numerical quintile between 1 and 5.
Within the print set, the 1st quintile contains the 20% of slides most likely to contain
abnormalities, the 2nd quintile contains the 20% of slides next most likely to contain
abnormalities, and so on, with the 5th quintile containing the 20% of the slides least likely
to contain abnormalities.
•
Each slide receives an individual rank from 1 to n, where a rank of 1 indicates a slide
most likely to contain abnormality and
n
is the slide least likely to contain abnormality
(
n
is the number of
Review
slides in a print set, minus the scant cellularity slides). For
example, the second slide listed on this report received an individual rank of 33/109, or
33 out of 109. For additional details, see Section 10.4.1.
•
The percentage value is the individual rank expressed as a percentage of the total of
n
.
7
Process Status
: A message appears in this area indicating the status of the scan. The
message most likely to be seen is
Scan
Completed
. However, when slides are rejected, or
the scanning process did not complete, a reason is printed under the Message column of the
report. See Section 21, for a list of these messages and the reasons they appear.
8
Process Date
: This area provides the date and the time the slide was processed.
Summary of Contents for FocalPoint GS
Page 10: ...BD FocalPoint GS Imaging System Instrument User s Manual 10...
Page 44: ...BD FocalPoint GS Imaging System Instrument User s Manual 44...
Page 54: ...BD FocalPoint GS Imaging System Instrument User s Manual 54...
Page 58: ...BD FocalPoint GS Imaging System Instrument User s Manual 58...
Page 76: ...BD FocalPoint GS Imaging System Instrument User s Manual 76...
Page 86: ...BD FocalPoint GS Imaging System Instrument User s Manual 86...
Page 110: ...BD FocalPoint GS Imaging System Instrument User s Manual 110...
Page 126: ...BD FocalPoint GS Imaging System Instrument User s Manual 126...
Page 156: ...BD FocalPoint GS Imaging System Instrument User s Manual 156...
Page 192: ...BD FocalPoint GS Imaging System Instrument User s Manual 192...
Page 200: ...BD FocalPoint GS Imaging System Instrument User s Manual 200...
Page 204: ...BD FocalPoint GS Imaging System Instrument User s Manual 204...
Page 206: ...BD FocalPoint GS Imaging System Instrument User s Manual 206...
Page 210: ...BD FocalPoint GS Imaging System Instrument User s Manual 210...
Page 212: ...BD FocalPoint GS Imaging System Instrument User s Manual 212...
Page 218: ...BD FocalPoint GS Imaging System Instrument User s Manual 218...
Page 224: ...BD FocalPoint GS Imaging System Instrument User s Manual 224...