BD FocalPoint™ GS Imaging System Instrument User’s Manual
98
10.4.3
Processing Status Information
Each slide result report confirms if the slide was completely and successfully imaged by the
BD FocalPoint™ Slide Profiler. If a slide was not successfully processed, a message stating the
reason for the unsuccessful scan accompanies the status.
Action
Scanning
Result
Adequacy*,
#
Squamous
Component
Endocervical
Component
Inflammation/
Obscuration
Review
—Instrument detected
potential cellular abnormalities. Slide
must be reviewed by a
Cytotechnologist.
Scan completed
SBLB,
Satisfactory
Detected
Detected/
Insufficient
50–75%
<50%
Review
—The instrument did not
detect adequate squamous
component. The slide must be
reviewed by a Cytotechnologist.
Scan completed
Unsatisfactory
Insufficient
Detected/
Insufficient
50–75%
<50%
Review
—The instrument detected
inflammation and/or obscuration. The
slide must be reviewed by a
Cytotechnologist.
Scan completed
Unsatisfactory
Detected/
Insufficient
Detected/
Insufficient
>75%
Review
—The instrument did not
detect adequate cellular material for
analysis. The slide must be reviewed
by a Cytotechnologist.
Scant cellularity
Unsatisfactory
Insufficient
Process Review
—The instrument
did not complete scanning or
rejected the slide for the reason
indicated. In some cases, the
problem can be corrected and the
slide can be rerun. For problems that
cannot be corrected and rerun, or
upon rerun, still do not complete
scanning, those slides must be
reviewed by a Cytotechnologist.
–Slide not
scanned
–Slide not
detected
–Unable to
analyze slide
–Staining not
within limits
–Image quality
not within
limits
–Slide
preparation
type mismatch
Rerun
—The instrument was not
functioning within limits when the
slide was processed. The slide must
be rerun when the instrument is
functioning properly, or these slides
must be reviewed by a
Cytotechnologist.
Tray rejected
Rerun
—The operator ejected the
tray before the slide could complete
processing or the system
experienced a power failure or
another type of interruption. Such
slides must be either rerun or
reviewed by a Cytotechnologist.
–Slide not
scanned
–User
requested halt
to slide
processing
–Scanning
incomplete
Empty
—The system did not detect a
slide in this tray position. If a slide is
present and was not detected, the
slide must be rerun, or these slides
must be reviewed by a
Cytotechnologist.
Slide not
detected
* Adequacy is reported only for conventional slides
#
Squamous component, endocervical component and inflammation/obscuration are reported for both BD SurePath™
and Conventional slides.
Table 10
-
1 – Possible Slide Results
Summary of Contents for FocalPoint GS
Page 10: ...BD FocalPoint GS Imaging System Instrument User s Manual 10...
Page 44: ...BD FocalPoint GS Imaging System Instrument User s Manual 44...
Page 54: ...BD FocalPoint GS Imaging System Instrument User s Manual 54...
Page 58: ...BD FocalPoint GS Imaging System Instrument User s Manual 58...
Page 76: ...BD FocalPoint GS Imaging System Instrument User s Manual 76...
Page 86: ...BD FocalPoint GS Imaging System Instrument User s Manual 86...
Page 110: ...BD FocalPoint GS Imaging System Instrument User s Manual 110...
Page 126: ...BD FocalPoint GS Imaging System Instrument User s Manual 126...
Page 156: ...BD FocalPoint GS Imaging System Instrument User s Manual 156...
Page 192: ...BD FocalPoint GS Imaging System Instrument User s Manual 192...
Page 200: ...BD FocalPoint GS Imaging System Instrument User s Manual 200...
Page 204: ...BD FocalPoint GS Imaging System Instrument User s Manual 204...
Page 206: ...BD FocalPoint GS Imaging System Instrument User s Manual 206...
Page 210: ...BD FocalPoint GS Imaging System Instrument User s Manual 210...
Page 212: ...BD FocalPoint GS Imaging System Instrument User s Manual 212...
Page 218: ...BD FocalPoint GS Imaging System Instrument User s Manual 218...
Page 224: ...BD FocalPoint GS Imaging System Instrument User s Manual 224...