BD FocalPoint™ GS Imaging System Instrument User’s Manual
96
10.3.2
Slides Not Successfully Processed
If the system cannot process a slide because of physical limitations, a mechanical failure of the
device, or an empty slot in a tray, one of the following actions is assigned to the slide:
Process
Review
,
Rerun
, or
Empty
. When a slide triggers one of these process failure actions, if the
failure is not resolved by rerunning the slide, then that slide must be manually screened.
Additionally, for slides with an action of either
Process Review
or
Rerun
that are manually
screened and then signed out as NILM, 10% must be randomly selected for QC review. Each of
the process failure actions is described in the following paragraphs.
Process Review Slides
Slides receiving an action of
Process Review
could not be analyzed because of slide physical
characteristics or specimen limitations.
Some
Process Review
slides with physical damage can
be repaired and rerun on the BD FocalPoint™ Slide Profiler. If a slide cannot be rerun, it must be
manually screened.
Rerun Slides
A slide receives a result of
Rerun
when processing results are not available. Slides receive this
result under the following circumstances:
•
during processing, the system falls outside specified operating limits.
•
during processing, a tray is manually ejected.
•
slide processing is interrupted (for example, power failure).
Rerun
slides should be reprocessed by the BD FocalPoint™ instrument. If the slides cannot be
successfully reprocessed, they must be manually screened.
Empty
Slide tray positions receiving an action of
Empty
indicate that the device did not detect a slide in
the tray position and the slide position is marked as empty on the results reports.
10.4 Slide Result Report Options
After the BD FocalPoint™ Slide Profiler processes and classifies slides, various reports that
show the slide processing results can be printed. Each time reports are printed, a print set, which
consists of all the slide results accumulated since the last time reports were printed, is created.
These slide result reports provide a variety of slide results information, which is explained in the
following subsections.
At the end of this section, Table 10
1 lists all the possible slide results generated by the
BD FocalPoint™ Slide Profiler, including the action required for each result, scanning result,
specimen adequacy information, and information on the presence or absence of endocervical
component. This information also appears on many of the slide result reports. For a complete
discussion of each type of report, see Section 12.
Summary of Contents for FocalPoint GS
Page 10: ...BD FocalPoint GS Imaging System Instrument User s Manual 10...
Page 44: ...BD FocalPoint GS Imaging System Instrument User s Manual 44...
Page 54: ...BD FocalPoint GS Imaging System Instrument User s Manual 54...
Page 58: ...BD FocalPoint GS Imaging System Instrument User s Manual 58...
Page 76: ...BD FocalPoint GS Imaging System Instrument User s Manual 76...
Page 86: ...BD FocalPoint GS Imaging System Instrument User s Manual 86...
Page 110: ...BD FocalPoint GS Imaging System Instrument User s Manual 110...
Page 126: ...BD FocalPoint GS Imaging System Instrument User s Manual 126...
Page 156: ...BD FocalPoint GS Imaging System Instrument User s Manual 156...
Page 192: ...BD FocalPoint GS Imaging System Instrument User s Manual 192...
Page 200: ...BD FocalPoint GS Imaging System Instrument User s Manual 200...
Page 204: ...BD FocalPoint GS Imaging System Instrument User s Manual 204...
Page 206: ...BD FocalPoint GS Imaging System Instrument User s Manual 206...
Page 210: ...BD FocalPoint GS Imaging System Instrument User s Manual 210...
Page 212: ...BD FocalPoint GS Imaging System Instrument User s Manual 212...
Page 218: ...BD FocalPoint GS Imaging System Instrument User s Manual 218...
Page 224: ...BD FocalPoint GS Imaging System Instrument User s Manual 224...